| Customization: | Available |
|---|---|
| After-sales Service: | Good |
| Warranty: | 1 Year |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Temperature Fluctuation | ±0.5℃ |
Humidity Fluctuation | ≤2.5% |
Temperature Deviation | ≤±2.0℃ |
Humidity Deviation | ≤±3% |
Heating Rate | -70℃to+150℃ ≥5℃/min (Average over the entire range) |
Cooling Rate | +150℃to-70℃ ≥5℃/min (Average over the entire range) |
Load Condition | None |
Operating Noise | A-weighted sound level ≤70 dB(A) |
(Measurement taken in a soundproof room with minimal echo at an ambient temperature of 25℃. | |
Using A-weighting, the average value is measured at 9 points. | |
Each test point is 1 meter horizontally from the noise source and 1 meter above the ground.) | |
Thermal Insulation Structure | 1.Outer wall material: A3 steel plate with spray coating. |
2.Inner wall material: Brushed stainless steel plate SUS#304. | |
3.Thermal insulation material for the box body: 100mm high-temperature-resistant rigid polyurethane foam. | |
4.Thermal insulation material for the door: 100mm high-temperature-resistant rigid polyurethane foam. | |
Bottom Structural Strength | The bottom of the test chamber is designed with track load-bearing capacity: ≤100kg/m² (load). |
Air Conditioning Channel | 1.Stainless steel long-shaft centrifugal fan: 2 units, 90W. |
2. Fan, heater, evaporator (also dehumidifier), drainage device, pressure balance port, adjustable air guide plate, temperature sensor. | |
Standard Configuration of the Test Chamber | 1.Window: 400x600x40mm, 3 layers of vacuum tempered glass. |
2.Embedded flat handle. | |
3.Door hinge: SUS #304 imported hinge. | |
4.Internal energy-saving light: LED energy-saving light. | |
5.Lead hole: φ50mm, 1 unit, located on the right side. | |
Door | 1.Single door, opens outward, hinge on the left, handle on the right (when facing the front of the chamber). |
2.Equipped with a safety door lock mechanism (the door can be opened from inside the test chamber). The door is equipped with electric anti-condensation heating, and the hollow glass observation window has a viewing area of approximately (W400×H600mm). The door frame is equipped with anti-condensation electric heating. | |
Control Panel | Controller with 7-inch display, temperature (humidity) control display, operation indicator light, and lighting button. |
Mechanical Room | Refrigeration unit, compressor water tray, pressure relief device, heating device. |
Power Distribution Cabinet | Power distribution board, cooling fan, main power switch, transformer, intermediate relay, time relay, solid-state relay, AC contactor, circuit breaker. |
Heater | 1.High-quality custom heating wire set. |
2.Heating control mode: SSR (solid-state relay), non-contact periodic pulse width modulation. | |
3.Heating power: approximately 4KW. | |
Power Line Hole | Located on the back of the chamber. |