EN12600 Building Glass Pendulum Impact Test Device
1,Application:
Pendulum impact tester is use for the construction of glass, safety glass impact test, is to test the safety of glass in the same mass under the impact of different height under the impact of different penetration resistance or strength of a special device, its technology in line with GB9962, GB9963 standard The relevant provisions.
Production standards: GB9656, GB9962, GB15763, GB / T5137, GB / T9963, GB14681, GB17841, GB18045, GB / T11944 and other standards.
2,Technology Parameter:
Impact mode | Pendulum style |
Swing arm radius | About 1960mm |
Impact height | 190~1200mm(The height of the impactor relative to the specimen center line) adjustable |
Impactor weight | 45kg±0.1kg |
Lift control deviation | 0-30mm |
Sample size | About H1938xW876±2mm |
Rubber strip size | W20±2mm thickness:10±1mm hardness:60±5IRHD |
Sample Wooden frame internal dimensions | H1910xW847±5mm |
Outer fame size | About W1066×D2760×H3100mm |
Weight | About 750kg |
Power | 1∮,AC220V,5A |
Wire rope diameter | 3mm±0.1mm |
Wire rope length | ≥1524mm |
Maximum impact height | 1200mm/2200mm |
Impactor trip Method | Pneumatic Column Traction Tripping |
Height measuring element | Own height pole |
Fixed way | Anchor bolts fixed |
Frame material | Steel |
3,Operation method:
1Remove the wooden frame clamp and remove the front wooden frame. Clean the front of the wooden frame.
2 upright into the test glass (high 1938±2mmX wide and 876±2mm), the bottom of the test glass placed along the bottom of the wood, the glass sample is located in the middle of the wooden frame and contact with the rubber strip.
3. Before loading the wood frame, the front wood frame rubber strips should be pressed against the test glass.
4. clamping the wood frame with a fixture, clamping force should be uniform.
5. in the console up or down to adjust the height, the standard impact height corresponding to 190mm, 450mm, 600mm, 750mm, 900mm, 1200mm.
6. impactor ear hanging into the release device pin, press the console up button to raise the shot pocket, to reach the test height, stop lifting.