Model BND-LD Tracking Index Test Apparatus
Product Description:
Introdution:Tracking Index Text Apparatus BND-LD creates moist conditions by providing drops between electrodes to measure the electrical breakdown(tracking) properties of an insulating material .
Standard:IEC60112 ,IEC60745, IEC60335-1, IEC60598-1.
Application: It is widely used for electrical safety assessment of electrical apparatus.This machine mainly tests the resistance tacking permance of solid insulating materials,which may be exposed to moisture and dirty environment.
Features:automation control systems;digital display.
Instruction: The Tracking Tester is a fully automatic instrument contained in its own cabinet to maximize the safety of the operator. Inside the well lit cabinet, a test specimen is placed on a glass plate and two platinum electrodes are lowered onto the top surface. A gauge block is supplied with the instrument to ensure the electrodes have the proper distance between them. The table height is easily adjustable to ensure the proper angle of the electrodes and height of drops. The rest of the test is controlled from outside the test chamber with the door closed. The voltage, limiting current, drip rate and number of drips are all set on the control panel. An operator should observe the test to see if a flame ignites. The instrument sounds an alarm if there is a tracking failure of at least two seconds.
Product show
Detailed Information
the electrode material | rectangular electrode - platinum metal |
the electrode pressure | 1.0 ± 0.05N |
the distance between electrodes | 4.0mm ± 0.01mm, angle 60 ° ± 5 |
the electrode voltage | 100 ~ 600V (48 ~ 60HZ) adjustable short-circuit current 1.0 ± 0.1A, voltage drop is less than 10% |
drops of liquid height | 30 ~ 40mm adjustable |
the size of the drops | 44 to 55 drops / 1cm3 Adjustable |
drops Time | 30s ± 5s, |
the number of drops | 0 to 9999 times adjustable |
the leakage judge | loop current greater than 0.5A and maintain the 2 seconds the relay to cut off the current, indicating that the sample failed |
the test process | automatic control of test procedures, independent ventilation |
reference standards | IEC60112 ,IEC60745, IEC60335-1, IEC60598-1 |
the studio volume | 0.5 cubic |